Beckman Coulter AU5800 Chemistry Analyzer Recall for 47 Units Over Calibration Delay Issues (2026)
Beckman Coulter Mishima K.K. recalled 47 AU5800 chemistry analyzers sold nationwide in the United States after a software delay can cause calibration and reagent databases to fail to open. The issue occurs when a calibration monitor using Microsoft SQL Server 2022 Express remains open for more than 15 minutes, potentially halting analysis. Healthcare facilities should stop using the devices and (a