Beckman Coulter Mishima K.K. Recalls

2 recalls found for Beckman Coulter Mishima K.K.. Check if any of your products are affected.

HIGHFDA DEVICE

CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment

Beckman Coulter Mishima K.K.
A delay
Read more
HIGHFDA DEVICE

DxC 700 AU, REF: B86444, B86446

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.

Beckman Coulter Mishima K.K.
A delay
Read more