HarmonyAIR A-Series Surgical Lighting System
The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
Brand hub
The the plus and minus buttons (+/-) used to control light intensity on the light heads may detach.
There is a potential for paint delamination (chipping/peeling) on the lighting system's light heads.
Wire connected to the electrical box may shift out of its intended position, which may result in electrical arcing. Electrical arcing remains internal within the unit (no risk of shock to users).
Steris recalled the AMSCO 7052HP Washer/Disinfector on November 19, 2025, after reports of electrical arcing. The issue results from a wire potentially shifting out of position, posing a fire risk. No injuries or shocks have been reported.
This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.