HIGHFDA DEVICE

Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers:...

Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.

Bayer Medical CareHealth & Personal CareMedical DevicesCatalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 876290076072945886566648

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
April 24, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
April 24, 2026
Hazard Level
HIGH
Brand
Bayer Medical Care
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
Bayer Medical Care
Model numbers
Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007, 60729458, 86566648, Affected batch numbers: 251702, 252302, 252502, 252602, 252702 +9 more
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    April 24, 2026

  2. Reported by FDA DEVICE

    June 10, 2026

  3. RecallRadar source check

    June 16, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Bayer Medical Care, Inc. or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for Bayer Medical Care recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Avanta Multi-Patient Administration Tube Sets (AVA 500 MPAT (Part 2 of 2)) used with the Medrad¿ Avanta Injection System; Catalog number: AVA 500 MPAT (Part 2 of 2); Affected material numbers: 87629007, 60729458, 86566648;. Reason: Supplier changes resulted in a fit issue with the Contrast Flow Control Valve (CFCV) interface on the Avanta Injector System. Multiple customer reports indicate that the stopcock cannot be reliably engaged with the CFCV snap interface. In instances where engagement is achieved through applied force, the connection is unstable, and the stopcock may disengage before or during contrast injection, leading to procedure interruption.. Classification: Class II. Quantity: 11,300 units (1650 US, 9650 OUS). Distribution: Worldwide - US Nationwide distribution including in the states of AL, AR, AZ, CA, CO, DE, FL, IA, IL, IN, KS, KY, LA, MA, MD, ME, MO, NC, NY, OH, OK, PA, SC, TN, TX, VA, WI and the countries of AE, BA, BE, CA, CH, CN, DE, DK, ES, FR, GB, HK, HR, IL, KW, MT, MX, NL, RO, RU, SE, ZA.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
Catalog number: AVA 500 MPAT (Part 2 of 2) Affected material numbers: 87629007
60729458
86566648
Affected batch numbers: 251702
252302
+12 more
Affected States
ALL
Report Date
June 10, 2026
Recall Status