Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

Hamilton Medical AGHealth & Personal CareMedical DevicesREF/UDI-DI: 260127/0763000280295617630002802953260128/07630002802963

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
May 29, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
May 29, 2026
Hazard Level
HIGH
Brand
Hamilton Medical AG
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
Hamilton Medical AG
Model numbers
REF/UDI-DI: 260127/07630002802956, 17630002802953, 260128/07630002802963, 17630002802960, 260167/07630002802970, 17630002802977, 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    May 29, 2026

  2. Reported by FDA DEVICE

    July 8, 2026

  3. RecallRadar source check

    July 14, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Hamilton Medical AG or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for Hamilton Medical AG recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.. Reason: Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement. Classification: Class I. Quantity: 27,270 pcs. Distribution: US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
REF/UDI-DI: 260127/07630002802956
17630002802953
260128/07630002802963
17630002802960
260167/07630002802970
+3 more
Affected States
ALL
Report Date
July 8, 2026
Recall Status
ACTIVE