Quick Facts at a Glance
- Recall Date
- May 29, 2026
- Hazard Level
- HIGH
- Brand
- Hamilton Medical AG
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Hamilton Medical AG
- Model numbers
- REF/UDI-DI: 260127/07630002802956, 17630002802953, 260128/07630002802963, 17630002802960, 260167/07630002802970, 17630002802977, 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
May 29, 2026
Reported by FDA DEVICE
July 8, 2026
RecallRadar source check
July 14, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Hamilton Medical AG or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for Hamilton Medical AG recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.. Reason: Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement. Classification: Class I. Quantity: 27,270 pcs. Distribution: US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.