HIGH

ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for...

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

Quick Facts at a Glance

Recall Date
March 31, 2026
Hazard Level
HIGH
Brand
Erbe Medical
Geographic Scope
1 states

Hazard Information

Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Erbe Medical, LLC or your healthcare provider for instructions

Get instant alerts for Erbe Medical recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

ERBEFLO CleverCap CO2: Hybrid CO2 Tubing/Cap Set for Olympus Scopes & CO2 Sources, REF: 20325-239; Hybrid CO2 Tubing/Cap Set for Olympus Scopes & UCR, REF: 20325-240; Hybrid CO2 Tubing/Cap Set for Olympus Scopes, REF: 20325-206; Hybrid CO2 Tubing/Cap Set for Pentax Scopes, REF: 20325-207; ERBEFLO CleverCap Hybrid CO2 Tubing/Cap Set for Fujifilm Scopes, REF: 20325-248; Hybrid CO2 Tubing/Cap Set for Fujinon Scopes, REF: 20325-208. Reason: Devices providing sterile water/CO2/air to endoscopes were manufactured with a distal irrigation segment connector configuration that may be more susceptible to unintended water flow from distal tip under unsupported third-party connector use conditions, which may lead to water aspiration, which may progress to aspiration pneumonia, acute respiratory distress syndrome, or respiratory failure.. Classification: Class II. Quantity: 260,962. Distribution: US Nationwide distribution including in the states of IL, NY, CA, WA, UT, KY, PA, TX, AL, OH, FL, IN, NJ, NH, MI, CT, AR, MN, WI, AZ, CO, WV, MT, HI, TN, GA, VA, OK, RI, MO, NM, ID, SC, WY, KS, NC, MS, OR, MD, MA, DE, NE, AK, IA, DC, ME, LA, SD, ND, VT, PR, NV.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
REF/UDI-DI/Lots: 20325-239/04065655000375/W4464506
W4464507
W4464512
W4464514
W4464515
+15 more
Affected States
ALL
Report Date
May 13, 2026
Recall Status
ACTIVE