HIGHFDA DEVICE

Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

Baxter HealthcareHealth & Personal CareMedical DevicesUDI: 00887761981492Lot Numbers: All lot numbers distributed beginning on 4/28/2025

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
May 21, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
May 21, 2026
Hazard Level
HIGH
Brand
Baxter Healthcare
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
Baxter Healthcare
Model numbers
UDI: 00887761981492, Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    May 21, 2026

  2. Reported by FDA DEVICE

    July 8, 2026

  3. RecallRadar source check

    July 14, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Baxter Healthcare Corporation or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for Baxter Healthcare recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Hillrom VOLARA System P.CIRCUIT KIT AC (Single Patient Use Circuit), Model/Catalog Number REF M08473; Oscillation Lung Expansion Therapy, positive pressure breathing device. Reason: There have been reports of air and medication leakage from the nebulizer cup during therapy of Volara system patient circuits. The leakage can lead to an oxygen level drop in the patient and ineffective nebulization, which may affect the delivery of the prescribed treatment. The issue has been associated with improper locking of the nebulizer cup after adding medication during user setup.. Classification: Class I. Quantity: 16187 units. Distribution: Worldwide distribution - US Nationwide and the country of Canada.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
UDI: 00887761981492
Lot Numbers: All lot numbers distributed beginning on 4/28/2025
Affected States
ALL
Report Date
July 8, 2026
Recall Status
ACTIVE

Related Recalls

Health & Personal Care
HIGH

Baxter Infusion Pump Recalled Due to Incomplete Testing

Baxter Healthcare recalled five infusion pumps on January 13, 2026. The pumps were released without full testing, including critical occlusion alarm checks. Healthcare providers and patients must stop using the device immediately.

Baxter Healthcare
Pumps were
Read more