Quick Facts at a Glance
- Recall Date
- June 8, 2026
- Hazard Level
- HIGH
- Brand
- Medline Industries, LP
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Medline Industries, LP
- Model numbers
- DYKMBNDL116G UDI-DI 10198459156991 Lots 25GBG086 25JBF391 25KBA114 25KBN017 25LBQ601
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
June 8, 2026
Reported by FDA DEVICE
July 29, 2026
RecallRadar source check
August 4, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Medline Industries, LP or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for Medline Industries, LP recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Medline convenience kits containing Bard¿ Silastic¿ Foley Catheters (French sizes 16, 18, 20) 5cc balloons. KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116G. Reason: Affected kits contain recalled Bard Silastic Foley Catheters. Catheters were recalled by manufacturer due to increased amount of raw material present on the surface of the catheter. Issue may result in the use of urinary catheters that exhibit staining on the catheter surface; raw material present in increased amounts may be harmful.. Classification: Class II. Quantity: 980 kits. Distribution: US Nationwide distribution in the states of CA, FL, MD. PA, and TX.
Browse More Recalls
Safety Guide
Not sure what to do next? Our guide walks you through the process step by step.
Read: How to Get Refunds and ReplacementsWant to Know First?
Get instant alerts for recalls that affect you. Free forever.