HIGHFDA DEVICE

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

Medline Industries, LPHealth & Personal CareMedical DevicesUDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
May 28, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
May 28, 2026
Hazard Level
HIGH
Brand
Medline Industries, LP
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
Medline Industries, LP
Model numbers
UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    May 28, 2026

  2. Reported by FDA DEVICE

    July 8, 2026

  3. RecallRadar source check

    July 14, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Medline Industries, LP or your healthcare provider for instructions. Notification method: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get instant alerts for Medline Industries, LP recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B. Reason: Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.. Classification: Class II. Quantity: 9,272 kits. Distribution: US Nationwide distribution in the states of AZ, CA, FL, IL, MO, NJ, OH, OK, TN, TX, WI.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720
Affected States
ALL
Report Date
July 8, 2026