Quick Facts at a Glance
- Recall Date
- June 16, 2026
- Hazard Level
- HIGH
- Brand
- Olympus Corporation of the Americas
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Olympus Corporation of the Americas
- Model numbers
- Model Number: UHI-4. UDI: 04953170324147, 04953170467219. All Serial Numbers.
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
June 16, 2026
Reported by FDA DEVICE
July 29, 2026
RecallRadar source check
August 4, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Olympus Corporation of the Americas or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for Olympus Corporation of the Americas recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Olympus High Flow Insufflation Unit - UHI-4. Model Number: UHI-4. Reason: Potential for unexpected instances where the front panel display turns off unexpectedly, and air delivery (insufflation) ceases.. Classification: Class II. Quantity: 3,087 units. Distribution: Worldwide distribution - US Nationwide and the countries of Australia, Bolivia, Brazil, Canada, Chile, China, Colombia, Costa Rica, Ecuador, El Salvador, Germany, Honduras, Hong Kong, India, Ireland, Japan, Korea, Malaysia, Mexico, Myanmar, New Zealand, Paraguay, Peru, Russia, Singapore, Sri Lanka, Taiwan, Thailand, United Kingdom, Uruguay, Vietnam.