Optima XR240amx, Mobile Digital Radiographic System

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

GE Medical SystemsHealth & Personal CareMedical DevicesUDI-DI: 00840682122351 and 00840682123686System IDs: 252847PP249281MSL2402

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
June 5, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
June 5, 2026
Hazard Level
HIGH
Brand
GE Medical Systems
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
GE Medical Systems
Model numbers
UDI-DI: 00840682122351 and 00840682123686, System IDs: 252847PP249, 281MSL2402, 291OPTI1, 312CCHSUPER7, 318487OPTIMA1, 360744OPT2402, 501686OPTI1 +12 more
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    June 5, 2026

  2. Reported by FDA DEVICE

    July 22, 2026

  3. RecallRadar source check

    July 28, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact GE Medical Systems, LLC or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for GE Medical Systems recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Optima XR240amx, Mobile Digital Radiographic System. Reason: GE HealthCare has become aware that in certain Optima XR200/220/240amx systems that have had a column replacement completed, functional testing of the safety pawl may not have been performed to confirm proper assembly. If a failure of the internal counterpoise cable occurs, the Column Safety mechanism will not work, causing the arm with X-ray tube and collimator to drop downward on the column track. This could result in patient injury during an exam.. Classification: Class II. Quantity: 24 units. Distribution: Worldwide - US Nationwide and the countries of Brazil, Canada, Korea, Mexico, Saudi Arabia, Spain, UK.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
UDI-DI: 00840682122351 and 00840682123686
System IDs: 252847PP249
281MSL2402
291OPTI1
312CCHSUPER7
+15 more
Affected States
ALL
Report Date
July 22, 2026
Recall Status