Quick Facts at a Glance
- Recall Date
- May 26, 2026
- Hazard Level
- HIGH
- Brand
- Philips North America
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Philips North America
- Model numbers
- UDI/DI 00884838000407, Serial Numbers: DE79569306, DE79569332, DE79569360, DE79569417, DE79569442, DE79569482, DE79569307 +12 more
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
May 26, 2026
Reported by FDA DEVICE
July 1, 2026
RecallRadar source check
July 7, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Philips North America Llc or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for Philips North America recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Philips Avalon Fetal Monitor, FM20, Part number M2702A, Part M2703A. Reason: Monitor has incorrect assembly of the speaker connector which can cause device cables to be pulled out of the housing and increase the risk of intermittent or permanent loss of speaker output.. Classification: Class II. Quantity: 285 units. Distribution: Worldwide distribution - US Nationwide and The countries of AE, AT, AU, BE, BG, BH, BR, CA, CH, CN, CR, CZ, DE, ES, FI, FR, GB, GR, HK, ID, IL, IN, IT, JO, JP, KR, KR, LT, LU, LV, MX, MY, NL, NZ, OM, SA, SE, SG, SI, TR, TW, ZA.
Browse More Recalls
Safety Guide
Not sure what to do next? Our guide walks you through the process step by step.
Read: How to Get Refunds and ReplacementsWant to Know First?
Get instant alerts for recalls that affect you. Free forever.