Quick Facts at a Glance
- Recall Date
- June 25, 2026
- Hazard Level
- HIGH
- Brand
- ResMed
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- ResMed
- Model numbers
- Catalog Number: Astral 100 UDI-DI code: 00619498003259 Serial Numbers: Astral Ventilators-Serial number < 22241890149 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070 Catalog Number: Astral 150 UDI-DI code: 00619498270033 Serial Numbers: Astral Ventilator-Serial numbers < 22241440282 PCBA Spare Parts via physical PCBA or Ventilator UI - Characters 2 8 < 2707658 PCBA Spare Parts in Box - Box serial number < 22241978070
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
June 25, 2026
Reported by FDA DEVICE
August 5, 2026
RecallRadar source check
August 12, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact ResMed Ltd. or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for ResMed recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Resmed Astral 100/150 Resmed Astral 100/50 is a ventilator system that provides continuous or intermittent ventilatory support for patients weighing more than 11lb (5kg) who require mechanical ventilation.. Reason: Due to a Supercapacitor electrolyte leakage on Printed Circuit Board Assembly (PCBA) may damage the charger chip, causing fail-safe state and interruption of ventilation therapy.. Classification: Class I. Quantity: 168,069 devices globally. Distribution: Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belgium, Bhutan, Brazil, Bulgaria, Canada, Chile, Colombia, Croatia, Cyprus, Czech Republic, Denmark, Ecuador, Egypt, Estonia, Finland, France, French Guiana, French Polynesia, Germany, Great Britain, Greece, Guadeloupe, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Islamic Republic of Iran, Israel, Italy, Japan, Jordan, Kenya, Republic of Korea, Kuwait, Latvia, Lebanon, Lithuania, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Saint Martin (French part), Saudi Arabia, Singapore, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, United Arab Emirates, Uruguay, and Viet Nam.
Browse More Recalls
Safety Guide
Not sure what to do next? Our guide walks you through the process step by step.
Read: How to Get Refunds and ReplacementsWant to Know First?
Get instant alerts for recalls that affect you. Free forever.