electronic Recalls

325 recalls tagged with “electronic”.

Health & Personal Care
HIGH
FDA DEVICE

SEDECAL SA Mobile X-ray System SM_40HF_BDC3 Recalled for Cleaning Risk (1 Unit, 2025)

SEDECAL SA recalled 1 unit of SM_40HF_BDC3 Mobile X-ray System distributed in California, Illinois and New Jersey. The device is not water-resistant and must be cleaned strictly per the manual. Healthcare facilities should stop using the device immediately and follow the manufacturer’s recall instructions.

SEDECAL SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalls 22 SM-32HF-Batt Mobile X-ray Systems for Cleaning Risk (2025)

Sedecal SA recalled 22 SM-32HF-Batt 32KW analog mobile X-ray systems distributed to California, Illinois and New Jersey. The units are not water-resistant and must be cleaned strictly per the manual, with explicit consequences for improper cleaning. Healthcare facilities should stop using the devices immediately and follow the manufacturer's recall instructions sent by email.

Sedecal SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalls 3-Unit Mobile X-ray System 40KWFXPLUS.004 Over Cleaning and Water-Resistance Conc

Sedecal SA recalled 3 units of the Mobile X-ray system 40KWFXPLUS.004 distributed in California, Illinois and New Jersey after warning users the device is not water-resistant and must be cleaned strictly per the manual. Improper cleaning could have consequences. Stop using the device and contact Sedecal SA or your healthcare provider for instructions.

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal Mobile X-ray System Recall for 1 Unit Over Water Exposure Cleaning Hazard (2025)

Sedecal SA recalls 1 unit of Model 40KWMOBILE3.899 Mobile X-ray system distributed to consignees in the United States, including California, Illinois and New Jersey. The device is not water-resistant and must be cleaned strictly according to the manual, with the notice outlining the potential consequences of improper cleaning. Healthcare providers and patients should stop using the device immediat

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SM-40HF-B-D-C Mobile X-ray System Recall for Cleaning Risk (2025)

Sedecal recalled 2 units of the SM-40HF-B-D-C 40KW 55G/60G Mobile X-ray system after warning consignees that the device is not water-resistant and must be cleaned exactly according to the manual. The notice outlines the potential consequences of improper cleaning. Healthcare facilities in California, Illinois and New Jersey should stop using the device immediately and contact Sedecal for further 1

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SM-40HF-B-D-C 40KW Mobile X-ray System Recall 2025

Sedecal SA recalled 4 units of the SM-40HF-B-D-C 40KW Mobile X-ray System distributed to U.S. facilities in California, Illinois and New Jersey. The equipment is not water-resistant and must be cleaned strictly according to the manual. Facilities should stop using the device immediately and follow the manufacturer’s recall instructions.

Sedecal SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SM-40HF-B-D-C 40KW Mobile X-ray System Recalled (2025)

Sedecal SA recalled a single SM-40HF-B-D-C 40KW mobile X-ray unit distributed to California, Illinois and New Jersey. The device is not water-resistant and must be cleaned per the manual, with explicit warnings about improper cleaning. Stop using the device and contact Sedecal SA or your healthcare provider for instructions.

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalls 1 Unit 40KW-FX.899 Mobile X-ray System Over Cleaning Hazard (2025)

Sedecal SA recalled 1 unit of the 40KW-FX.899 Mobile X-ray System distributed to CA, IL and NJ. The device is not water-resistant and must be cleaned strictly per the manual. Improper cleaning could cause adverse effects. Stop using the device and contact Sedecal SA or a healthcare provider for instructions.

Sedecal SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal 40KWFX Mobile X-ray System Recall Affects 55 US Units (2025)

Sedecal recalled 55 Mobile X-ray System units sold to facilities in California, Illinois and New Jersey. The devices are not water-resistant and must be cleaned exactly as described in the manual, with clear warnings about the consequences of improper cleaning. Hospitals and healthcare providers should stop using the device immediately and follow Sedecal's recall instructions.

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalls Shimadzu RF.004 Mobile X-ray System in 2025

Sedecal SA is recalling the Shimadzu RF.004 Mobile X-ray System after warnings that the device is not water-resistant and must be cleaned strictly according to the manual. Hospitals in California, Illinois, and New Jersey are affected. Stop using the device and follow recall instructions.

Sedecal SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalls 181 SM-40HF-B-D-C Mobile X-ray Systems for Cleaning Not Water-Resistant Risk (202

Sedecal SA recalls 181 SM-40HF-B-D-C mobile X-ray systems in 2025 after concerns about cleaning and water resistance. The device is not water-resistant and must be cleaned strictly according to the manual. Manufacturers warn of potential consequences if cleaning is not followed. Hospitals and clinics should stop using the device and follow recall instructions immediately.

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA SM-40HF-B-D-C Mobile X-ray System Recalled for Cleaning Not-Water-Resistant Risk (142 US)

Sedecal SA recalled 142 SM-40HF-B-D-C 40KW mobile X-ray systems sold to consignees in California, Illinois and New Jersey. Sedecal warned the units are not water-resistant and must be cleaned exactly as the manual prescribes, outlining the consequences of improper cleaning. Hospitals and clinics should stop using the devices immediately and follow the manufacturer's recall instructions.

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SM-32HF-B-D-C 32KW Mobile X-ray System Recall Expanded (2025)

Sedecal SA recalled 8 SM-32HF-B-D-C 32KW 50G Mobile X-ray systems sold in the US and Canada after notifying consignees by email that the equipment is not water-resistant and must be cleaned strictly per the manual. The notice outlines possible consequences of improper cleaning. Hospitals should stop using the devices immediately and follow the recall instructions.

SEDECAL SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal MobileDiagnost wDR 2.2 Recall for 1,198 Units in 2025

Sedecal recalled 1,198 MobileDiagnost wDR 2.2 mobile X-ray systems distributed in California, Illinois and New Jersey. The company reminded users the equipment is not water-resistant and must be cleaned strictly according to the manual. Improper cleaning could have consequences outlined by Sedecal. Health care providers should stop using the device immediately and follow recall instructions.

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalls 9 SM-40HF-Batt Mobile X-ray Systems (2025)

Sedecal SA is recalling 9 units of the SM-40HF-Batt 40KW analog Mobile X-ray System distributed in California, Illinois and New Jersey. The company sent an email to consignees warning the equipment is not water-resistant and must be cleaned strictly per the manual. Improper cleaning could cause system damage. Hospitals and clinics should stop using the devices immediately and contact Sedecal SA or

Sedecal SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA 40KWFXPLUS.002 Mobile X-ray System Recalled in 2025 (8 Units)

Sedecal SA recalled 8 mobile X-ray units in the United States after notifying consignees that the equipment is not water-resistant and must be cleaned strictly per the manual. The notice warns that improper cleaning could have consequences described in the manual. Healthcare providers and patients should stop using the devices immediately and contact Sedecal SA or their healthcare provider for the

Sedecal
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalled 5 SM-40HF-B-D-C Mobile X-ray Systems for Cleaning Warning (2025)

Sedecal SA recalled 5 SM-40HF-B-D-C 40KW 55G Mobile X-ray systems distributed in the US (CA, IL, NJ) after warning that the equipment is not water-resistant and must be cleaned strictly per the manual. The notice outlines the possible consequences of improper cleaning. Healthcare facilities should stop using the devices immediately and contact Sedecal SA for instructions.

Sedecal SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalled 5 SM-40HF-B-D-C Mobile X-ray Systems Over Cleaning Risk (2025)

Sedecal SA recalled 5 SM-40HF-B-D-C Mobile X-ray systems distributed in California, Illinois and New Jersey after notifying consignees via email. The notice says the equipment is not water-resistant and must be cleaned strictly per the manual. Hospitals and clinics should stop using the devices immediately and follow the recall instructions.

SEDECAL SA
Sedecal sent
Read more
Health & Personal Care
HIGH
FDA DEVICE

Sedecal SA Recalled 1 Mobile X-ray System 40KWFXPLUS.889 in 2025 for Cleaning Not Water-Resistant

Sedecal SA recalled 1 unit of the Mobile X-ray System 40KWFXPLUS.889 distributed in California, Illinois and New Jersey. The recall concerns cleaning practices and the device not being water-resistant. Hospitals and clinics should stop using the device immediately and follow recall instructions. Contact Sedecal SA for instructions via the recall notice distributed by email.

Sedecal SA
Sedecal sent
Read more