Ascend Laboratories Recalls ARIPIPRAZOLE 10 mg Tablets Over Superpotent Drug Risks (2025)
ARIPIPRAZOLE
Superpotent drug
These recalls involve products that pose serious risks of injury or death. Check if you own any of these products and take immediate action if needed.
Immediate attention



The product in incorrectly labeled. The label indicates that the skive hole should be above the balloon, while it is actually positioned below, and vice versa.
a potential product mix where the package labeled as Catalog Number 0580-1-442, lot A00976, may contain Catalog Number 0580-1-352, lot G8754849, and vice versa. Patient labels may also be impacted.
Siemens Medical Solutions USA recalled 71 MAGNETOM Lumina MRI systems worldwide, including US and 50+ countries, due to ice blockage in the magnet venting system that could cause a helium leak during a quench.