Quick Facts at a Glance
- Recall Date
- April 20, 2026
- Hazard Level
- HIGH
- Brand
- Abiomed
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Abiomed
- Model numbers
- Product Code: 0042-0000-US, UDI-DI: 00813502010022, Serial Numbers: IC2449, IC2451, IC2452, IC2453, IC2459, IC2460 +12 more
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
April 20, 2026
Reported by FDA DEVICE
June 3, 2026
RecallRadar source check
June 10, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Abiomed, Inc. or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for Abiomed recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Automated Impella Controller (AIC); Product Code: 0042-0000-US;. Reason: A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.. Classification: Class I. Quantity: 91 units. Distribution: Worldwide - US Nationwide distribution in the states of CA, CT, FL, LA, MA, MI, MO, MS, ND, NY, OH, PA, TX, WY and the countries of Brazil, Canada, Colombia, Hong Kong, Italy, Kuwait, Saudi Arabia, UAE, United Kingdom;
Browse More Recalls
Safety Guide
Not sure what to do next? Our guide walks you through the process step by step.
Read: How to Get Refunds and ReplacementsWant to Know First?
Get instant alerts for recalls that affect you. Free forever.