Quick Facts at a Glance
- Recall Date
- April 2, 2026
- Hazard Level
- HIGH
- Brand
- Siemens Healthcare Diagnostics
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Siemens Healthcare Diagnostics
- Model numbers
- Siemens Material Number (SMN): 10736515, UDI-DI: 00809708121860, Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only)
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
April 2, 2026
Reported by FDA DEVICE
May 6, 2026
RecallRadar source check
May 13, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Siemens Healthcare Diagnostics Inc or your healthcare provider for instructions. Notification method: Letter
Get instant alerts for Siemens Healthcare Diagnostics recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;. Reason: Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.. Classification: Class II. Quantity: 71725 (38025 US, 33700 OUS). Distribution: Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.