epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;

Siemens Healthcare DiagnosticsHealth & Personal CareMedical DevicesSiemens Material Number (SMN): 10736515UDI-DI: 00809708121860Lot numbers: 01-26024-70

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
April 2, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
April 2, 2026
Hazard Level
HIGH
Brand
Siemens Healthcare Diagnostics
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
Siemens Healthcare Diagnostics
Model numbers
Siemens Material Number (SMN): 10736515, UDI-DI: 00809708121860, Lot numbers: 01-26024-70, 01-26026-70, 01-26028-70, 01-26030-70 (OUS only), 02-26027-80, 02-26028-80 (OUS only)
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    April 2, 2026

  2. Reported by FDA DEVICE

    May 6, 2026

  3. RecallRadar source check

    May 13, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Siemens Healthcare Diagnostics Inc or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for Siemens Healthcare Diagnostics recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

epoc BGEM BUN Test Card (25 pk) for the epoc Blood Analysis System; Siemens Material Number (SMN): 10736515;. Reason: Use of the affected product can result in discrepant low pH and measured Total Carbon Dioxide (mTCO2) results and corresponding calculated values in arterial, venous, and capillary patient samples. An erroneously low pH or mTCO2 reading can lead to inappropriate treatment of nonexistent acidosis (respiratory of metabolic) or missed diagnosis of alkalosis (a true high pH: respiratory or metabolic alkalosis), potentially causing harm through unnecessary interventions with the former or failure to deliver appropriate treatment with the latter.. Classification: Class II. Quantity: 71725 (38025 US, 33700 OUS). Distribution: Worldwide distribution - US Nationwide and the countries of Italy, Spain, Austria, Portugal, Ireland, France, Romania, Canada, India, Philippines, Mexico.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
Siemens Material Number (SMN): 10736515
UDI-DI: 00809708121860
Lot numbers: 01-26024-70
01-26026-70
01-26028-70
+3 more
Affected States
ALL
Report Date
May 6, 2026