HIGHFDA DEVICE

Heraeus, PALAMIX uno. Material Number: 66057893.

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Heraeus Medical GmbHHealth & Personal CareMedical DevicesMaterial Number: 66057893UDI-DI: 04260102133587Batch: all batches with numbers below 223060

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
March 25, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
March 25, 2026
Hazard Level
HIGH
Brand
Heraeus Medical GmbH
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
Heraeus Medical GmbH
Model numbers
Material Number: 66057893, UDI-DI: 04260102133587, Batch: all batches with numbers below 223060
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    March 25, 2026

  2. Reported by FDA DEVICE

    May 6, 2026

  3. RecallRadar source check

    May 13, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Heraeus Medical GmbH (Dental Division) or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for Heraeus Medical GmbH recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Heraeus, PALAMIX uno. Material Number: 66057893.. Reason: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).. Classification: Class III. Quantity: 51,771 units. Distribution: Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
Material Number: 66057893
UDI-DI: 04260102133587
Batch: all batches with numbers below 223060
Affected States
ALL
Report Date
May 6, 2026
Recall Status
ACTIVE

Related Recalls

Health & Personal Care
HIGH

Heraeus Medical Recalls PALACOS Bone Cement Due to Ampoule Breakage

Heraeus Medical GmbH recalled PALACOS R pro 40 and R pro 80 bone cement on December 9, 2025. The recall follows multiple complaints of ampoule breakage affecting product usability. Healthcare providers must stop using the product immediately.

Heraeus Medical GmbH
increase of
Read more