HIGH

Heraeus, PALAMIX uno. Material Number: 66057893.

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Quick Facts at a Glance

Recall Date
March 25, 2026
Hazard Level
HIGH
Brand
Heraeus Medical GmbH
Geographic Scope
1 states

Hazard Information

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Heraeus Medical GmbH (Dental Division) or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for Heraeus Medical GmbH recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Heraeus, PALAMIX uno. Material Number: 66057893.. Reason: Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).. Classification: Class III. Quantity: 51,771 units. Distribution: Worldwide distribution - US Nationwide and the countries of Aland Islands, Algeria, Australia, Austria, Belgium, Brazil , Chile, China, Cyprus, Czech Republic, Denmark, Ecuador, Finland, France, Germany, Greece, Hungary, India, Ireland, Israel, Italy, Kosovo, Lithuania, Luxembourg, Malaysia, Malta, Mexico, Netherlands, Nigeria, Norway, Peru, Philippines, Poland, Portugal, Puerto Rico, Rep. of Korea, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Thailand, United Kingdom, Uruguay, Utd.Arab Emir.;

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
Material Number: 66057893
UDI-DI: 04260102133587
Batch: all batches with numbers below 223060
Affected States
ALL
Report Date
May 6, 2026
Recall Status
ACTIVE

Related Recalls

HIGH

Heraeus, PALAMIX duo. Material Number: 66057897.

Based on long-term stability testing, it was determined that in units older than 1.5 years (17 months) the mixing rod may, in rare cases, come loose from its lock during the mixing process. Notification is to reduce the shelf life from 3 years to 1.5 years (17 months).

Heraeus Medical GmbH
Based on
Read more
Health & Personal Care
HIGH

Heraeus Medical Recalls PALACOS Bone Cement Due to Ampoule Breakage Hazard

Heraeus Medical GmbH recalled PALACOS R+G pro 40 and 80 bone cements on December 9, 2025. The recall follows multiple reports of ampoule breakage that prevent proper use of the product. Patients and healthcare providers must stop using the affected devices immediately.

Heraeus Medical GmbH
increase of
Read more
Health & Personal Care
HIGH

Heraeus Medical Recalls PALACOS Bone Cement Due to Breakage Risk

Heraeus Medical GmbH recalled PALACOS MV+G pro 40 and MV+G pro 80 bone cement on December 9, 2025. The recall follows complaints about ampoule breakage, which can prevent proper cement formation. This issue could jeopardize patient safety during medical procedures.

Heraeus Medical GmbH
increase of
Read more
Health & Personal Care
HIGH

Heraeus Medical Recalls PALACOS Bone Cement Due to Ampoule Breakage

Heraeus Medical GmbH recalled PALACOS R pro 40 and R pro 80 bone cement on December 9, 2025. The recall follows multiple complaints of ampoule breakage affecting product usability. Healthcare providers must stop using the product immediately.

Heraeus Medical GmbH
increase of
Read more