HIGH

Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

Quick Facts at a Glance

Recall Date
March 12, 2026
Hazard Level
HIGH
Brand
Insulet
Geographic Scope
1 states

Hazard Information

Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Insulet Corporation or your healthcare provider for instructions. Notification method: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get instant alerts for Insulet recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Omnipod 5 Pods. Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7. Reason: Due to a manufacturing defect, certain Omnipod 5 Pods from 49 lots have an internal soft cannula tear that results in insulin leaking into the Pod instead of being delivered to the user regardless of basal or bolus delivery. This defect results from damage to the unexposed portion of the soft cannula during manufacturing, which would result in a compromised fluid path. The primary failure mode is pump under-delivery due to loss of insulin to an internal leak; in some cases, the defect may also lead to pump shutoff and cessation of insulin delivery when leaked insulin contacts Pod circuity in a manner that results in an electrical short. Under-delivery of insulin (both basal and bolus insulin) or cessation of insulin put users at risk of hyperglycemia, and complications from acute and chronic hyperglycemia, including dehydration, blurry vision, nausea, vomiting, altered mental status, diabetic ketoacidosis (DKA), hyperglycemic hyperosmolar syndrome (HHS), or even death. Users may require hospitalization or medical intervention to treat severe adverse health consequences. Not all devices with the defect will issue an alarm or alert the user. If there is sufficient leakage of insulin to cause a short in the circuity, the Pod will issue a Hazard Alarm that stops all insulin delivery and alerts the user to replace their Pod. In addition, if a user s glucose is trending high and is not responding to insulin delivery, the system may reach the maximum amount of insulin microboluses allowed by the system and trigger the Automated Delivery Restriction (ADR) alert that tells users to check their blood glucose and take appropriate actions (i.e., ADR is a response to persistent hyperglycemia and maximum automated delivery constraints rather than a direct detection of the leak). The magnitude of under-delivery is unknown and based on multiple factors, including how much insulin is being delivered, whether an alarm and/or alert triggers, whether and when the user recognizes the device defect, the duration of Pod use, and the size of the tear.. Classification: Class I. Quantity: 1,240,115 units. Distribution: Nationwide distribution.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Brand
Model Numbers
Model/Catalog number: PT-001662 ASM 5PK Pod STRL OPS G6G7. UDI-DI: 10385083000527. Lot Codes: PH1U02252541
PH1U03282511
PH1U03282522
PH1U03312511
PH1U03312521
+15 more
Affected States
ALL
Report Date
April 22, 2026
Recall Status
ACTIVE

Related Recalls

Health & Personal Care
MEDIUM

Taro Pharmaceuticals Withdraws Diclofenac Sodium Gel Over Viscosity Issues

Taro Pharmaceuticals announced a market withdrawal of Diclofenac Sodium, Topical Gel, 3% on January 27, 2026. The withdrawal stems from out of specification viscosity results. Healthcare providers and consumers should stop using the product immediately.

DESMOPRESSIN ACETATE
Defective container:
Read more
Health & Personal Care
HIGH

K.C. Pharmaceuticals Recalls 8 OTC Eye Drops for Sterility Concerns (2026)

K.C. Pharmaceuticals recalled eight OTC eye drops sold nationwide under CVS Health, Walgreens and Kroger brands after sterility could not be verified. The recall covers all lots within expiry through Oct 31, 2026. Nonsterile eye drops can harbor bacteria and fungus, causing serious infections that may lead to vision loss. Stop using recalled products and return them for a full refund at the store.

K.C. Pharmaceuticals
Nonsterile eye
Read more
Health & Personal Care
HIGH

Biocon Pharma Prazosin Hydrochloride 1mg 100-Count Capsules Recalled for Nitrosamine Impurity (2026)

Biocon Pharma Inc. recalled 88,008 bottles of Prazosin Hydrochloride capsules, 1 mg, 100-count bottles, nationwide in the United States. A cGMP deviation detected N-nitroso-prazosin impurity C above acceptable limits. Consumers and healthcare providers should stop using this product immediately and contact Appco Pharma LLC or a healthcare provider for guidance.

Biocon Pharma
cGMP deviation:
Read more
Health & Personal Care
HIGH

Biocon Pharma Prazosin Hydrochloride 5mg Recall 28,157 Bottles Over Nitrosamine Impurity (2026)

Biocon Pharma Inc. recalled 28,157 bottles of PRAZOSIN HYDROCHLORIDE capsules, distributed nationwide in the USA. The recall cites cGMP deviation due to detection of N-nitroso-prazosin impurity C above acceptable limits. Stop using the product immediately and contact Appco Pharma LLC or your healthcare provider for guidance.

Prazosin Hydrochloride
cGMP deviation:
Read more
Health & Personal Care
HIGH

Biocon Prazosin Hydrochloride 2 mg Capsules Recalled for Nitrosamine Impurity (58,896 bottles, 2026)

Biocon Pharma Inc. recalled 58,896 bottles of Prazosin Hydrochloride Capsules, USP, 2 mg, 100-count bottle. The product was distributed nationwide in the USA and manufactured by Appco Pharma LLC. A cGMP deviation detected N-nitroso-prazosin impurity C above acceptable limits. Consumers and healthcare providers should stop using this product immediately and contact Appco Pharma LLC or their health‑

Prazosin Hydrochloride
cGMP deviation:
Read more