Quick Facts at a Glance
- Recall Date
- January 17, 2011
- Hazard Level
- HIGH
- Brand
- Abiomed
- Geographic Scope
- 1 states
What to Do Now
Use this page like a recall checklist: verify the product first, then act on the official remedy.
Immediate steps
- 1Stop using the product until you confirm whether it is included in the recall.
- 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
- 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
- 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.
Check these identifiers
- Brand
- Abiomed
- Model numbers
- 1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324, IC1346, IC1480, IC1481, IC1543, IC1544, IC1572, IC1595 +12 more
- Where affected
- ALL
Recall Timeline
Key dates and source checks for this recall record.
Recall announced
January 17, 2011
Reported by FDA DEVICE
June 3, 2026
RecallRadar source check
June 10, 2026
Consumer action
Use the official remedy and keep documentation.
Hazard Information
Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.
What You Should Do
Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Abiomed, Inc. or your healthcare provider for instructions. Notification method: Other
Get instant alerts for Abiomed recalls
Be the first to know. Free instant alerts to your inbox.
Full Description
Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.. Reason: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.. Classification: Class I. Quantity: 625 units. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Denmark, France, Germany, Italy, Kuwait, Netherlands, Norway, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.