HIGH

Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller,...

Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

Quick Facts at a Glance

Recall Date
January 17, 2011
Hazard Level
HIGH
Brand
Abiomed
Geographic Scope
1 states

Hazard Information

Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Abiomed, Inc. or your healthcare provider for instructions. Notification method: Other

Get instant alerts for Abiomed recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Automated Impella Controller (AIC) with the below product descriptions and corresponding Product Codes. 1. Impella Controller, Packaged, CA; Product Code: 0042-0000-CA. 2. Impella Controller, Packaged, EU; Product Code: 0042-0000-EU. 3. Impella Controller, Packaged, UK; Product Code: 0042-0000-UK. 4. Impella Controller, Packaged, US; Product Code: 0042-0000-US.. Reason: Retrospective submission for following issues identified: 1. Alarm failures, power-path faults, electrical-short risks, and inaccurate motor-current sensing were identified in certain AIC consoles. 2. There is improper alignment between the purge cassette and the motor drive in the AIC. This can result in a piston block event and lead to the inability to complete priming of the purge system, requiring the user to switch to backup AIC console.. Classification: Class I. Quantity: 625 units. Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Denmark, France, Germany, Italy, Kuwait, Netherlands, Norway, Saudi Arabia, Spain, Switzerland, United Arab Emirates, United Kingdom.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Brand
Model Numbers
1. Product Code: 0042-0000-CA. GTIN: 00813502011272. Serial Numbers: IC1324
IC1346
IC1480
IC1481
IC1543
+15 more
Affected States
ALL
Report Date
June 3, 2026
Recall Status
ACTIVE

Related Recalls

HIGH

Automated Impella Controller (AIC); Product Code: 0042-0000-US;

A retrospective review identified that there are AIC units requiring specific hardware updates to address potential safety concerns. These hardware updates have been implemented through Abiomed s servicing process; therefore, AICs that have not received service are pending these updates. Issue 1: The proximity of the internal Video Graphics Array cable to the Digital Signal Processor chipset on the Impellatronic printed circuit assembly could potentially result in Electrostatic Discharge coupling into the Digital Signal Processor which may interrupt motor controls. Issue 2: There is a potential for the Compact Flash Memory Card dislodgement due to external applied forces, which could result in AIC startup failures and data logging issues. Issue 3: Improper routing of the fan wire within the AIC could lead to fan wire damage, which may result in console boot-up failure. Issue 4: Potential capacitor related issues on the Power Battery Manager, including the potential for the capacitors to cause pump stop, purge stop, and/or single fan fuse failures.

Abiomed
A retrospective
Read more
Health & Personal Care
HIGH

Abiomed Purge Cassette Recall 2026: 33,107 Impella Pump Sets Worldwide Over Purge Leaks

Abiomed has issued a Class I recall for 33,107 Purge Cassette units used with Impella pump sets worldwide due to purge leaks in Generation 1 cassettes. The leaks raise the risk of compromised device performance. Patients and healthcare providers should stop using the affected devices immediately and follow Abiomed's recall instructions.

Abiomed
Increased risk
Read more
Health & Personal Care
HIGH

Abiomed Impella Automated Controller Recall 62 Units in 2026

Abiomed recalled 62 Automated Impella Controller units sold to hospitals internationally. The recall warns of a delayed Purge System Blocked alarm display when used with first generation Impella 5.5 pumps. Hospitals should stop using the device immediately and contact Abiomed for instructions.

Abiomed
Potential that
Read more
Health & Personal Care
HIGH

Abiomed Impella RP with SmartAssist Recall Affects 291 Units Worldwide in 2026

Abiomed recalled 291 Impella RP with SmartAssist devices sold worldwide, including the US. A differential pressure sensor may malfunction and cause readings to drift. Patients and healthcare providers should stop using the device immediately and follow the manufacturer’s recall instructions.

Abiomed
Differential pressure
Read more
Health & Personal Care
HIGH

Abiomed Impella RP Recall 179 Units for Sensor Drift Risk (2026)

Abiomed recalled 179 Impella RP devices worldwide after a differential pressure sensor may drift. The fault can cause sensor values to drift, potentially affecting patient care. Stop using the device immediately and follow the recall instructions, including contacting Abiomed or your healthcare provider.

Abiomed
Differential pressure
Read more
Health & Personal Care
HIGH

Abiomed Recalls Impella 5.5 Devices Due to Packaging Error

Abiomed recalled two units of the Impella 5.5 with SmartAssist on December 4, 2025. The devices were packaged in incorrect outer cartons, posing a potential hazard. Patients and healthcare providers must stop using the devices immediately.

Abiomed
Device packaged
Read more
Health & Personal Care
HIGH

Abiomed Impella Controller Recall Aims to Address Cybersecurity Vulnerabilities Across 32 Product SK

Abiomed recalled 10,153 Automated Impella Controllers and related components due to potential cybersecurity vulnerabilities in the operating system. The recall spans multiple international product codes and was first issued for devices distributed nationwide and internationally. Providers should stop using the affected controllers and follow manufacturer instructions immediately.

Abiomed
Potential cybersecurity
Read more