Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067,...

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.Health & Personal CareMedical Devices1. Azurion 3 M12Model: 722063UDI: (01)00884838085275(21)122

Source-backed recall notice

No product image was available from the source feed.

Official source
FDA DEVICE
Recall date
April 30, 2026
Status
ACTIVE

Quick Facts at a Glance

Recall Date
April 30, 2026
Hazard Level
HIGH
Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Geographic Scope
1 states

What to Do Now

Use this page like a recall checklist: verify the product first, then act on the official remedy.

Immediate steps

  1. 1Stop using the product until you confirm whether it is included in the recall.
  2. 2Compare the product label, model number, UPC, color, size, and purchase location against the identifiers below.
  3. 3Follow the official remedy from FDA - Medical Devices; save photos, receipts, labels, and correspondence before contacting the company.
  4. 4If the product could harm a child, older adult, pet, or patient, move it out of reach immediately.

Check these identifiers

Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Model numbers
1. Azurion 3 M12, Model: 722063, UDI: (01)00884838085275(21)122, (01)00884838085275(21)64, (01)00884838085275(21)56, (01)00884838085275(21)131, (01)00884838085275(21)84, (01)00884838085275(21)132 +12 more
Where affected
ALL

Recall Timeline

Key dates and source checks for this recall record.

  1. Recall announced

    April 30, 2026

  2. Reported by FDA DEVICE

    June 3, 2026

  3. RecallRadar source check

    June 10, 2026

  4. Consumer action

    Use the official remedy and keep documentation.

Hazard Information

It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063, 722221), Azurion 3 M15 (722064, 722222), Azurion 5 M12 (722227), Azurion 5 M20 (722228), Azurion 7 B12 (722067, 722225), Azurion 7 B20 (722068, 722226), Azurion 7 M12 (722078, 722223), Azurion 7 M20 (722079, 722224).. Reason: It was identified that HDDs used in the PCs of Azurion and Allura systems may show a decrease in performance as they age, particularly beyond six years of service. Issues with an HDD may, depending on the specific system PC in which the affected HDD is installed, result in loss of imaging functionality. Additionally, in Allura systems, this may lead to loss of motorized movement, or loss of data. In some cases, a system restart may temporarily restore functionality.. Classification: Class II. Quantity: 3069 units (872 US, 2224 OUS). Distribution: Worldwide distribution - US Nationwide and the countries of Canada, Afghanistan, Albania, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussal, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Greece, Guinea, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Korea, South, Kosovo, Kuwait, Kyrgyzstan, Lao, Latvia, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Martinique, Mauritius, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Fed., Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Vietnam, Yemen.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
1. Azurion 3 M12
Model: 722063
UDI: (01)00884838085275(21)122
(01)00884838085275(21)64
(01)00884838085275(21)56
+15 more
Affected States
ALL
Report Date
June 3, 2026
Recall Status
ACTIVE

Related Recalls