Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA...

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

Quick Facts at a Glance

Recall Date
March 19, 2026
Hazard Level
HIGH
Brand
PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Geographic Scope
1 states

Hazard Information

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. or your healthcare provider for instructions. Notification method: Letter

Get instant alerts for PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA data.. Reason: Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.. Classification: Class II. Quantity: 1,033 units. Distribution: Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bnis insted Us Instas instes, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Estonia, Germany, Greece, Indonesia, Iraq, Italy, Japan, Kuwait, Latvia, Macao, Maldives, Mongolia, Morocco, New Zealand, Nicaragua, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Switzerland, TURKEY, Ukraine, Utd.Arab.Emir., Uzbekistan, Vietnam.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
Model Number: 001015. Software Version Number R1.x. UDI Number: (01)00884838099272(21)153
(01)00884838099272(21)67
(01)00884838099272(21)675
(01)00884838085367(21)178
(01)00884838099272(21)112
+15 more
Affected States
ALL
Report Date
April 22, 2026
Recall Status
ACTIVE

Related Recalls

Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 7 M12 Due to Hazardous Movement

Philips Medical Systems recalled 1,979 Azurion 7 M12 Systems on January 26, 2026. The table may move unexpectedly when the Reset Geometry button is pressed, posing a high risk of injury. Healthcare providers and patients should stop using the devices immediately.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 7 M20 Over Hazardous Table Movement

Philips Medical Systems recalled 5,136 Azurion 7 M20 systems on January 26, 2026. The devices may move unexpectedly when the Reset Geometry button is pressed, posing a danger to patients and healthcare providers. Consumers should stop using the devices immediately and contact the manufacturer for instructions.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 7 B20 System Over Movement Risk

Philips Medical Systems recalled 1,503 Azurion 7 B20 systems on January 26, 2026. The devices may move unexpectedly when the Reset Geometry button is pressed, posing a risk of injury. Healthcare providers must stop using the device and follow recall instructions immediately.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 5 M20 Table Over Movement Hazard

Philips Medical Systems Nederland B.V. recalled 922 Azurion 5 M20 systems on January 26, 2026, due to a defect that causes unexpected table movement. This hazard can occur even with an active table lock when the Reset Geometry button is pressed. The recall includes 101 units in the U.S. and 821 units distributed worldwide.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 3 M12 Due to Hazard

Philips Medical Systems recalled 291 Azurion 3 M12 systems on January 26, 2026, due to a risk of unexpected table movement. This may occur when the Reset Geometry button is pressed, even if a table lock is active. The recall affects 27 units in the U.S. and 264 units internationally.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 7 B12 Table Due to Safety Risk

Philips Medical Systems Netherlands B.V. recalled 655 Azurion 7 B12 systems on January 26, 2026. The recall affects models 722067, 722225, and 722235 due to unexpected table movement when the Reset Geometry button is pressed. This poses a high risk of injury to patients and healthcare providers.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 5 M12 Over Safety Risk

Philips Medical Systems recalled 326 Azurion 5 M12 systems on January 26, 2026. The table may move unexpectedly when the Reset Geometry button is pressed, posing a risk to patients and healthcare providers. Immediate cessation of use is advised.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more
Health & Personal Care
HIGH

Philips Medical Systems Recalls Azurion 3 M15 System Due to Hazard

Philips Medical Systems Nederland B.V. recalled 802 Azurion 3 M15 systems on January 26, 2026. The systems may move unexpectedly, posing a risk to patients and healthcare providers. Consumers should stop using the device immediately and follow recall instructions.

PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
Under certain
Read more