HIGH

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINAL FUSION, Medline Kit Number/SKU CDS860017AG

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

Quick Facts at a Glance

Recall Date
March 19, 2026
Hazard Level
HIGH
Brand
Medline Industries, LP

Hazard Information

During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.

What You Should Do

Patients and healthcare providers should stop using this device immediately. Follow the recall instructions provided by the manufacturer. Contact Medline Industries, LP or your healthcare provider for instructions. Notification method: Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit

Get instant alerts for Medline Industries, LP recalls

Be the first to know. Free instant alerts to your inbox.

No spamUnsubscribe anytime

Full Description

Medline medical procedure kits, containing Medline Neuro Sponges, labeled as follows: SPINAL FUSION, Medline Kit Number/SKU CDS860017AG. Reason: During an internal review, Medline identified higher-than-expected endotoxin levels on a representative Medline neuro sponge product, indicating that in-market product may be out-of-specification for endotoxin.. Classification: Class II. Quantity: 120 kits. Distribution: Worldwide distribution.

Safety Guide

Not sure what to do next? Our guide walks you through the process step by step.

Read: How to Get Refunds and Replacements

Want to Know First?

Get instant alerts for recalls that affect you. Free forever.

Product Classification

Product Details

Model Numbers
UDI/DI each 10198459237317
UDI/DI case 40198459237318: Lot Numbers: 25LBI308
25JBH227
25HBQ371
25HBP069
+6 more
Report Date
May 6, 2026
Recall Status
ACTIVE